MRI COMPARISON CHART, Philips Ingenia 3T vs. Toshiba Vantage Titan 3T

Posted on: 10/23/2012 3:54:16 PM under MRI Scanners

Philips Ingenia 3T vs. Toshiba Vantage Titan 3T

MRI COMPARISON CHART

Philips Ingenia 3T vs. Toshiba Vantage Titan 3T

MRI Model

> Ingenia 3.0T

> Vantage Titan 3T

Company

Philips Healthcare

Toshiba

FDA Clearance

YES, 2011

YES, 2011

Clinical Application

Whole Body multi-transmit

MRI

>> MAGNET

Configuration

Cylindrical

Short magnet, wide bore

Type

Superconducting

OR-96 SMT

Field Strength

3 T

3 T

Homogeneity
(35 or 40cm DSV), Vrms

<0.4 ppm @ 40cm DSV

< 1.4 ppm @ 40 cm DSV

5-Gauss Fringe Field
Radial Axial, m

3.05 4.95

5 gauss: 2.6 4.6

Cryogen Refill Period

Zero boil off, 5-7 years

Zero boil off magnet

Finished Gantry Weight, kg
(Covered)

5,800

6,375 with cryogens

Finished (Covered) Gantry Dimension, (L x W x H)

1.50 x 1.88 x 2.29 m

188 x 224 x 222 cm

>>PATIENT COMFORT

Minimum Finished Bore L-R Diameter (Closed Magnet)

70 cm

163 x 69 cm

Minimum A-P Dimension w/ Table Inserted

62 cm

53.5 cm

Table Width
(moving portion)

NA

61.4 cm

Table Capacity

550 lbs

440 lbs

Table Vertical Travel
(Min height-scanning height)

Approximately 36 cm

42.9 – 84.5 cm

Table Lateral Movement Range

NA

205 cm

Patient Cooling Features

Low SAR, 6 airflow settings

Adj. ventilation w/
improved airflow

Patient Illumination Features

Yes, + Ambient ring

Advanced illumination

Patient Operator
Call & Intercom

YES

YES

>>GRADIENT

Maximum Amplitude
Single Axis, mTm

45

30 per axis

Maximum Slew Rate
Single Axis, Tms

200

203 per axis

Cooling System Type

Liquid water mix

Water

>>COMPUTER SYSTEM

CPU Type

Intel Quad Core 2.4 GHz

Dual Core Dual CPU 3.0 GHz

CPU Memory Size

8,000 MB

8 GB and up

Reconstruction Hardware

Intel Quad Core 2.4 GHz

Dual Core Dual CPU 2.93 GHz

Reconstruction Memory Size

8,000 MB

24 GB

Image Storage Memory Type

HD, DVD+RW, DICOM PACS

RAID 10

Image Storage Media
Image Capacity

150 GB HD, 4.7 GB DVD

1.3 TB

DICOM 3.0 Classes Supported

Enhanced MR standard

YES

Display Monitor Displayable Area (L x W), cm

52 x 33

55.6 x 36.1

Simultaneous Scan & Reconstruction

YES

YES

>>RF SYSTEM

Channels
(Min, Max Configuration)

Min > 52, Max > 1,000  

16 (32 optional)

Parallel Imaging Features
(name, imageK-space)

SENSE, ktBLAST
(k-space, image)

Speeder (image),
JET (radial k-space)

>>COILS
(StandardOptional, #of elements, RF channels, parallel imaging support)

Brain

Std, 15 chan

Atlas Speeder head 16 el
(32-el opt; WIP)
array, std, receive

Head

Std, 15 chan

QD head coil, std,
transmit & receive

Spine

Std, 44 chan

Atlas Speeder spine 40 el array, std, receive

Neck

Std, 15 chan

Atlas Speeder head coil w/ neck attachment is used

Shoulder

Opt, 8 chan

NA

Body Torso

Opt, 54 chan

Atlas Speeder body 16 el array, std, receive

Knee

Opt, 8 chan

Large knee Speeder 7 el array, opt, transmit & receive

Cardiac

Opt, 54 chan

Atlas Speeder body & spine coils are used

Breast

Opt, 7 or 16 chan

NA

Wrist

Opt, 8 chan

Wrist-hand Speeder 6 el array, opt, receive

Neurovascular

Std; 40 chan

Atlas Speeder head coil w/ a neck attachment is used

Peripheral Vascular

Opt, 54 chan

Atlas Speeder body & spine coils are used

TMJ

Std, 15 chan

Flex coils & coil holder, std, Speeder support

Others

Opt

Flex coils f100 f150, std,
Speeder support

>> IMAGING FEATURES

Noncontrast Angiography

TRANCE, bTRANCE

FBI, CIA, Time-Slip, TSA

Spectroscopy

2-D, 3-D SENSE, spectro

H-MRS single-voxel & CSI

Motion Compensating
Radial Techniques

MultiVane

JET

Brain Volume Imaging

3-D BrainView

mVox

Minimum Repetition Time, msec

TSE TR 17

1.4

Minimum Echo Time, msec

TSE TE 1.8

0.4

FOV (AP, RL, HF,
Min to Max), cm

55 x 55 x 50 (0.05-55)

50 x 50 x 45
(min 5mm, max 550mm)

>> POWER REQUIREMENTS

Voltage (line voltage, phases)

400 480 3-phase

400V, 3-phase

kVA

86

90

AC, BTUhr
(for entire system)

56,000 peak

Std depending on the site


The Bottom Line
: These MRI systems are the BIG BOYS and will provide such a great enhancement to any facility. Pricing for these units are similar, however, warranties can vary depending on your facility’s location and your provider’s desire to expand their territory. What needs to be considered is the availability of parts after the warranty expires. None the less, both of these 3T MRI systems are so similar in their basic operations; it will only come down to pricing and extended warranty support.

Posted by:
Nathan Welch
MRI Specialist
407.438.7847   
Nathan@AmberUSA.com

 

 